Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Product picture thermocouple sensor cable probe TH52

iTHERM CableLine TH52
Cable probe

Imperial TC thermometer with connecting cable for use in many process and laboratory applications

TH52
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
5-year warranty

Engineered for Confidence

Our Fundamental Selection instruments feature an extended 5-year warranty – a clear testament of our trust in their long-term reliability and performance.

Should a warranty case occur, you can count on a fast replacement process and minimal administrative effort – giving you peace of mind when it matters most and helping you keep your operations running smoothly.

Warranty terms apply

The extended warranty is subject to specific conditions. For full details, please refer to the locally applicable General Terms and Conditions (GTC), which can be found on the respective country page of our website.

Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Thermocouple

  • Characteristic / Application

    US style
    cable probe
    process connection as compression fitting

  • Thermowell / protection tube

    without (not intended to use with thermowell)

  • Insert / probe

    mineral insulated (MI), flexible

  • Outer diameter protection tube / Insert

    1/16'' (1,59 mm)
    1/8'' (3,18 mm)
    3/16'' (4,76 mm)
    1/4'' (6,35 mm)

  • Max. immersion length on request

    up to 96" (2438 mm)
    others on request

  • Material protection tube/ thermowell

    316
    Alloy 600

  • Optional coating

    Not applicable

  • Process connection

    compression fitting:
    NPT1/8"
    NPT1/4"

  • Tip shape

    straight

  • Surface roughness Ra

    nicht definiert

  • Operating temperature range

    Type T:
    -270 °C ...400 °C
    (-454 °F ...752 °F)
    Type J:
    -210 °C ...1.200 °C
    (-346 °F ...2.192 °F)
    Type E:
    -270 °C ...1.000 °C
    (-454 °F ...1.832 °F)
    Type K:
    -270 °C ...1.372 °C
    (-454 °F ...2.500 °F)
    Type N:
    -270 °C ... 1.300 °C
    (-454 °F ...2.372 °F)

  • Max. process pressure (static)

    at 20°C: 40 bar (580 psi)'
    depends on configuration

  • Accuracy

    Standard acc. to ASTM E-230
    Special acc. to ASTM E-230

  • Response time

    63% rt = 0.3 s

  • Integration head transmitter

    Not applicable